Hematoxylin #14166
Product Information
Product Usage Information
Hematoxylin is a blue nuclear counterstain for use after the completion of immunohistochemical assays. It yields crisp nuclear staining with superior contrast with SignalStain® DAB Substrate Kit #8059 and is also compatible with SignalStain® Mounting Medium #14177.
A) Solutions and Reagents
Supplied Reagents
1. Hematoxylin #14166
Additional Reagents (Not Supplied)
1. Acid Rinse Solution: To prepare 100ml, add 2ml of glacial acetic acid to 98 ml of dH20.
2. Bluing Solution: To prepare 100ml, add 1.5 ml of 30% ammonium hydroxide (NH4OH) to 98.5ml of 70% ethanol
3. SignalStain® Mounting Medium #14177
B) Nuclear Counterstaining
1. Rinse slides in tap water.
2. Completely submerse slides in Hematoxylin #14166 and incubate for 1-5 minutes.
3. Wash slides with tap water until rinse water is clear.
4. Immerse slides 10 times in acid rinse solution.
5. Rinse slides 10 times in tap water.
6. Immerse slides for 1 minute in bluing solution
7. Rinse slides 10 times in tap water.
8. Mount sections with SignalStain® Mounting Medium #14177.
Storage
Product Description
Limited Uses
Except as otherwise expressly agreed in a writing signed by a legally authorized representative of CST, the following terms apply to Products provided by CST, its affiliates or its distributors. Any Customer's terms and conditions that are in addition to, or different from, those contained herein, unless separately accepted in writing by a legally authorized representative of CST, are rejected and are of no force or effect.
Products are labeled with For Research Use Only or a similar labeling statement and have not been approved, cleared, or licensed by the FDA or other regulatory foreign or domestic entity, for any purpose. Customer shall not use any Product for any diagnostic or therapeutic purpose, or otherwise in any manner that conflicts with its labeling statement. Products sold or licensed by CST are provided for Customer as the end-user and solely for research and development uses. Any use of Product for diagnostic, prophylactic or therapeutic purposes, or any purchase of Product for resale (alone or as a component) or other commercial purpose, requires a separate license from CST. Customer shall (a) not sell, license, loan, donate or otherwise transfer or make available any Product to any third party, whether alone or in combination with other materials, or use the Products to manufacture any commercial products, (b) not copy, modify, reverse engineer, decompile, disassemble or otherwise attempt to discover the underlying structure or technology of the Products, or use the Products for the purpose of developing any products or services that would compete with CST products or services, (c) not alter or remove from the Products any trademarks, trade names, logos, patent or copyright notices or markings, (d) use the Products solely in accordance with CST Product Terms of Sale and any applicable documentation, and (e) comply with any license, terms of service or similar agreement with respect to any third party products or services used by Customer in connection with the Products.