When harvesting cross-linked chromatin from tissue samples, the yield of chromatin can vary significantly between tissue types. The table below provides ranges for expected total yield of chromatin and expected DNA concentration from 25 mg of tissue or 4 x 106 HeLa cells, as determined in Section IV of the protocols.
For optimal ChIP results, we recommend using 5 to 10 µg of cross-linked and fragmented chromatin per IP; therefore, some tissues may require harvesting more than 25 mg per each IP.
SimpleChIP® Kit | Enzymatic | Sonication | ||
---|---|---|---|---|
Tissue / Cell | Total Chromatin Yield | Expected DNA Concentrations | Total Chromatin Yield | Expected DNA Concentrations |
Tissue / Cell | Total Chromatin Yield | Expected DNA Concentrations | Total Chromatin Yield | Expected DNA Concentrations |
Spleen | 20–30 µg per 25 mg tissue | 200–300 µg/ml | NT | NT |
Liver | 10–15 µg per 25 mg tissue | 100–150 µg/ml | 10–15 µg per 25 mg tissue | 100–150 µg/ml |
Kidney | 8–10 µg per 25 mg tissue | 80–100 µg/ml | NT | NT |
Brain | 2–5 µg per 25 mg tissue | 20–50 µg/ml | 2–5 µg per 25 mg tissue |
20–50 µg/ml |
Heart | 2–5 µg per 25 mg tissue | 20–50 µg/ml | 1.5–2.5 µg per 25 mg tissue | 15-25 µg/ml |
HeLa | 10–15 µg per 4 x 106 cells | 100–150 µg/ml | 10–15 µg per 4 x 106 cells | 100–150 µg/ml |
NT = not tested
In the SimpleChIP® Enzymatic protocol, optimal conditions for the digestion of cross-linked chromatin DNA to 150–900 bp fragments is highly dependent on the ratio of micrococcal nuclease to the amount of tissue or number of cells used in the digest. Below is a protocol for determination of the optimal digestion conditions for a specific tissue or cell type.
In the SimpleChIP® sonication protocol, optimal conditions for the fragmentation of cross-linked chromatin DNA are highly dependent on the number of cells, volume of sample, length of sonication, and sonicator power setting used. For each sonication sample, we recommend using 100–150 mg of tissue or 1 x 107–2 x 107 cells per 1 ml ChIP Sonication Nuclear Lysis Buffer. Below is a protocol for determining the optimal sonication conditions for a specific tissue or cell type.Resuspend nuclear pellet in 200 μl of 1X ChIP buffer + PIC. Incubate on ice for 10 min.
NOTE: Optimal sonication conditions can vary with different sample types and fixation times. Use the minimal number of sonication cycles required to generate the desired length of chromatin fragments. Over-sonication, indicated by >80% of total DNA fragments being shorter than 500 bp, can result in excessive damage to the chromatin and lower immunoprecipitation efficiency.
Problem | Possible Causes | Recommendation |
---|---|---|
Problem | Possible Causes | Recommendation |
1. Concentration of the fragmented chromatin is too low. | Not enough cells or tissue were used for the chromatin preparation or cell/tissue lysis was incomplete. | If DNA concentration of the chromatin preparation is close to 50 μg/ml, add additional chromatin to each IP to give at least 5 μg/IP and continue with protocol. |
Count a separate plate of cells before cross-linking to determine an accurate cell number. | ||
Enzymatic: visualize cell nuclei under microscope before and after sonication to confirm complete lysis of nuclei. | ||
2. Chromatin is under-fragmented and fragments are too large. Large chromatin fragments can lead to increased background and lower resolution. | Cells may have been over-crosslinked and/or too much input material (cells/tissue) was processed. | Shorten the crosslinking time within 10–30 minute range and/or reduce the amount of cell/tissues per sonication. |
Enzymatic: increase the amount of Micrococcal nuclease to the chromatin digestion or perform a time course for enzymatic digestion. | ||
Sonication: conduct a sonication time course. | ||
3. Chromatin is over-fragmented. Digestion of chromatin to mono-nucleosome length DNA may diminish signal during PCR quantification, especially for amplicons greater than 150 bp in length. Over-sonication of chromatin may disrupt chromatin integrity and denature antibody epitopes. | Enzymatic: Not enough cells or too much Micrococcal nuclease added to digestion. | Enzymatic: weigh tissue or count a separate plate of cells prior to cross-kinking to determine accurate cell number. Add more tissue or cells, or less Micrococal nuclease to the chromatin digest. |
Sonication: Conditions are too harsh. | Sonication: conduct a sonication time course to find a minimum output/duration to achieve appropriate sonication. | |
4. No product or very little product in the input PCR reactions. | Not enough DNA added to the PCR reaction or conditions are not optimal. | Add more DNA to the PCR reaction or increase the number of amplification cycles. Optimize the PCR conditions for experimental primer set using purified DNA from cross-linked and fragmented chromatin. |
PCR amplified region may span nucleosome-free region. | Design a different primer set and decrease length of amplicon to less than 150 bp. | |
Not enough chromatin added to the IP or chromatin is over-fragmented. | For optimal ChIP results add 5–10 μg chromatin per IP. See recommendations for problems 1 and 3 above. | |
5. No product in the positive control Histone H3-IP RPL30 PCR reaction. | Not enough chromatin or antibody added to the IP reaction or IP incubation time is too short. | Be sure to add 5–10 μg of chromatin and 10 μl of antibody to each IP reaction and incubate with antibody over-night and an additional 2 hr after adding Protein G beads. |
Incomplete elution of chromatin from Protein G beads. | Elution of chromatin from Protein G beads is optimal at 65°C with frequent mixing to keep beads suspended in solution. | |
6. Quantity of product in the negative control Rabbit IgG-IP and positive control Histone H3-IP PCR reactions is equivalent. | Too much or not enough chromatin added to the IP reaction. Alternatively, too much antibody added to the IP reaction. | Add no more than 15 μg of chromatin and 10 μl of histone H3 antibody to each IP reaction. Reduce the amount of normal rabbit IgG to 1 μl per IP. |
Too much DNA added to the PCR reaction or too many cycles of amplification. | Add less DNA to the PCR reaction or decrease the number of PCR cycles. For accurate quantitation, it is critical that PCR products are analyzed within the linear amplification phase of PCR. | |
7. No product in the Experimental Antibody-IP PCR reaction. | Not enough DNA added to the PCR reaction. | Add more DNA to the PCR reaction or increase the number of amplification cycles. |
Not enough antibody added to the IP reaction. | Typically a range of 1–5 μg of antibody is added to the IP reaction; however, the exact amount depends greatly on the individual antibody. Increase the amount of antibody added to the IP. | |
Antibody does not work for IP. | Choose an alternate, ChIP-validated antibody. |
posted March 2008
revised May 2017
Support: 877-678-8324 [email protected] • Orders: 877-616-2355 [email protected] • Web: www.cellsignal.com