For best results:
- Maintain an RNase-free and sterile workspace.
- It is highly recommended to use filtered pipette tips.
- When starting section B. Blocking (Day 2) until the end of the protocol, it is crucial to perform
the centrifugations at 850 x g (not RPM).
- When starting section C. Immunostaining (Day 2) in the protocol, it is crucial to maintain the
cell pellet submerged (in ~40 µL) when removing and discarding the supernatant.
- Do not vacuum aspirate when discarding the supernatant in any step; always use a pipette.
- Make sure the reagents are completely thawed before use.
NOTES:
- All InTraSeq™ 3' Assay Kit components are validated by CST in the 10x Genomics Chromium Single Cell 3'
Kits with Feature Barcoding technology.
- SAFE STOP - This indicates a safe stopping
point in the protocol
if stopping is necessary.
- This protocol is designed for a single InTraSeq™ assay (one sample). If processing multiple samples, all
quantities should be adjusted proportionally.
Solutions and Reagents
Materials Included in Kit (Store at -20°C)
- InTraSeq™ Blocking Buffer A #10471
- InTraSeq™ Blocking Buffer B #31701
- InTraSeq™ Staining Buffer #46005
- InTraSeq™ Wash Buffer #59335
- InTraSeq™ Reducing Agent #63760
- InTraSeq™ RNase Inhibitor #79298
- Rabbit (DA1E) mAb IgG XP® Isotype Control (InTraSeq™ 3' Conjugate 3000) #81472
- Histone H3 (D1H2) XP® Rabbit mAb (InTraSeq™ 3' Conjugate 3002) #68984
Required Reagents Not Included
- InTraSeq™ 3' Conjugates/Antibody Cocktails
- Two sterile 40-micron Falcon cell strainers
- Vortex mixer
- 1.5 mL Sterile Polypropylene Microcentrifuge Tubes - LoBind
- 15 mL Conical Sterile Polypropylene Centrifuge Tubes (preferably LoBind)
- 50 mL Conical Sterile Polypropylene Centrifuge Tubes (preferably LoBind)
- Phosphate Buffered Saline (PBS-1X) pH7.2 (Sterile) #9872
- Methanol #13604
Before starting:
(Optional) When you receive the kit, thaw and pipette mix the InTraSeq™ Blocking Buffer A, then aliquot
870 µL into 8 separate 1.5 mL sterile microcentrifuge tubes.
A. Cell Fixation (Day 1)
NOTE: If performing an immunostaining step prior to this
protocol, we highly recommend beginning immunostaining
with at least 2 million cells to guarantee 1-5 million cells
in step 1 below.
- Count and transfer 1-5 million cells to a 15 mL tube.
NOTE: Do not proceed with the protocol with less than 1
million cells.
- Centrifuge the cells at 300 x g for 5 min at 4°C.
- Remove and discard the supernatant, then wash the cells with 10 mL of ice-cold 1X PBS.
- Centrifuge the cells at 300 x g for 5 min at 4°C.
- Remove and discard supernatant, without disturbing the cell pellet, then resuspend the cells in 0.5 mL
of ice-cold 1X PBS.
- Turn on the vortex mixer and keep it on the “On” switch (not the “Auto/Touch” mode) at low
speeds (~speed 1-4 depending on your mixer). The sample should NOT be vortexed, but it should
rather be constantly mixed. If the mixer is fast enough to vortex the sample, decrease
the speed accordingly.
NOTE: It is very important to drop the ice-cold methanol
slowly (>30 sec) in the step below.
- Place the 15 mL tube on the vortex mixer while it is on the “On” switch then slowly add 4.5 mL of
ice-cold methanol in a dropwise manner (>30 sec).
- Incubate the cells overnight in the -20°C freezer. (SAFE STOP - sample
can be stored in the -20°C freezer for up to 7 days)
B. Blocking (Day 2)
From this point forward, at every
centrifugation step, it is crucial to spin
down the cells at 850 x g. Decreasing
the spin down speed will result in cell
loss.
-
Thaw the following reagents at room temperature:
- Thaw 1 bottle of InTraSeq™ Wash Buffer.
- Thaw the InTraSeq™ Blocking Buffer B.
- Thaw and pipette mix the InTraSeq™ Blocking Buffer A.
- Thaw and vortex mix InTraSeq™ Reducing Agent until the precipitate disappears.
- Once thawed, place the reagents on ice.
-
Prepare the heat/cooled InTraSeq™ Blocking Buffer B:
- Aliquot 130 µL InTraSeq™ Blocking Buffer B into a sterile 1.5 mL microcentrifuge tube. Return
the stock InTraSeq™ Blocking Buffer B to the freezer.
- Heat the aliquoted tube at 95°C for 5 min.
- Spin the tube down briefly then place it on ice. Use this heat/cooled InTraSeq™ Blocking Buffer
B in the steps below.
-
Prepare the scBlock (see table below), pipette mix. Store on ice.
scBlock |
Volume |
InTraSeq™ Blocking Buffer A |
860 µL |
Heat/cooled InTraSeq™ Blocking Buffer B |
98 µL |
InTraSeq™ Reducing Agent |
40 µL |
InTraSeq™ RNase Inhibitor |
2 µL |
TOTAL Volume |
1,000 µL |
- Centrifuge the cells at 850 x g for 5 min at 4°C.
- Remove and discard the supernatant, then add the scBlock to the cells. Pipette mix to resuspend
the cells, then keep the cells on ice for at least 30 min.
-
During the scBlock incubation, prepare the Antibody Master Mix as follows:
- Determine the volume of Antibodies (Vol.
of Abs.) based
on the number of InTraSeq™ 3' Antibody Cocktails and Individual InTraSeq™ 3' Conjugates (Ind.
InTraSeq™ 3' Conj. or Ind. Conj.) being used (see table
below). Limit the total volume to 30 µL.
|
Volume (Cocktails only) |
Volume (Ind. Conj. only) |
Volume (Cocktails + Ind. Conj.) |
InTraSeq™ 3' Cocktail |
5 µL (per cocktail) |
N/A |
___ µL (5 µL per cocktail) |
Ind. InTraSeq™ 3' Conj. |
N/A |
3 µL (per Ind. Conj.) |
___ µL (3 µL per Ind. Conj.) |
Vol. of Abs.
|
___ µL |
___ µL |
___ µL |
- Prepare the Antibody Master Mix (see table below) using the Vol. of Abs. (max 30 µL)
calculated above. Gently pipette
mix to minimize bubble formation, then store on ice.
Antibody Master Mix |
Volume (Cocktails only) |
Volume (Ind. Conj. only) |
Volume (Cocktails + Ind. Conj.) |
Vol. of Abs.
|
___ µL |
___ µL |
___ µL |
InTraSeq™ Wash Buffer |
30 µL - Vol.
of Abs. (see row
above) |
InTraSeq™ Staining Buffer |
55 µL |
55 µL |
55 µL |
Heat/cooled InTraSeq™ Blocking Buffer B |
10 µL |
10 µL |
10 µL |
Rabbit (DA1E) mAb IgG XP® Isotype Control (InTraSeq™ 3' Conjugate
3000) |
3 µL |
3 µL |
3 µL |
Histone H3 (D1H2) XP® Rabbit mAb (InTraSeq™ 3' Conjugate 3002) |
3 µL |
3 µL |
3 µL |
InTraSeq™ RNase Inhibitor |
2 µL |
2 µL |
2 µL |
TOTAL Volume |
103 µL |
103 µL |
103 µL |
- After the 30 min scBlock incubation, add 2 mL InTraSeq™ Wash Buffer to the cells and
pipette mix.
- Filter the cells through a sterile 40-micron Falcon cell strainer into a 50 mL tube.
NOTE: To maximize cell recovery, apply gentle
pressure to the pipette tip against the strainer while filtering. (See image)
-
Transfer the flow through/filtered cells into a new 15 mL tube and immediately start section
C. Immunostaining (Day 2).
NOTE: Store the remaining InTraSeq™ Wash Buffer at
4°C for use on Day 3.
C. Immunostaining (Day 2)
From this point forward, whenever removing and discarding the supernatant, it is crucial to maintain the
cell
pellet submerged (in ~40 µL). Drying the cells will negatively impact the RNA signal and result in
cell loss.
- Centrifuge the cells at 850 x g for 5 min at 4°C.
- Remove and discard supernatant while retaining ~40 µL in the tube.
- Add 100 µL Antibody Master Mix to the cells. Pipette mix to resuspend the cells.
- Transfer to a sterile 1.5 mL LoBind microcentrifuge tube and incubate at 4°C overnight (for ~16 hr).
NOTES:
- Do not incubate the cells for longer than 20 hr.
- It is crucial to maintain the overnight incubation at 4°C.
D. Washes and Count (Day 3)
Before starting:
- Add 1 mL Modified InTraSeq™ Wash Buffer to the 1.5 mL microcentrifuge tube containing the cells.
Pipette mix then transfer the mixture to a 15 mL tube.
- Add 2 mL Modified InTraSeq™ Wash Buffer so the total volume is 3 mL.
- Centrifuge at 850 x g for 5 min at 4°C. Remove and discard supernatant while retaining
~40 µL in
the tube.
- Add 1 mL Modified InTraSeq™ Wash Buffer. Pipette mix to resuspend the cells, then add
another 3 mL Modified InTraSeq™ Wash Buffer so the total volume is 4 mL.
- Repeat steps “3” and “4”, then filter the 4 mL cell suspension through a sterile 40-micron
Falcon cell strainer into a 50 mL tube. Transfer the flow through/filtered cells into a new 15
mL tube.
NOTE: To maximize cell recovery, apply gentle
pressure to the pipette tip against the strainer while filtering. (See image)
- Centrifuge the cells at 850 x g for 5 min at 4°C. Remove and discard supernatant while
retaining
~40 µL in the tube.
- Add 100 µL Modified InTraSeq™ Wash Buffer to the cells. Pipette mix to resuspend the cells,
then keep the cells on ice.
NOTE: From this point forward, keep the cells in the
Modified InTraSeq™ Wash Buffer including any necessary dilutions. Using any other buffer
or solution will result in RNA degradation.
- Count the cells. Refer to the Cell Suspension Volume Calculator Table in the 10x Genomics Chromium
Single Cell 3' Kits with Feature Barcoding technology protocol to determine the desired cell
concentration.
IMPORTANT: Dilute the cells using ONLY the Modified InTraSeq™
Wash Buffer to reach the desired cell concentration.
NOTE: If cell clumps are observed after counting, additional
pipette mixing is recommended before proceeding to the single-cell experiment.
- Process the cells using the 10x Genomics Chromium Single Cell 3' Kits with Feature Barcoding technology.
For Research Use Only. Not for Use in Diagnostic Procedures.