Can I set my centrifuge to RPM rather than RCF for spinning my samples generated using the Cell Fractionation Kit #9038?

It is important to use the RCF (relative centrifugal force) settings on your centrifuge rather than the RPM (revolutions per minute) settings when using the Cell Fractionation Kit #9038 as RCF accounts for the size of the rotor and is therefore a more accurate representation of the force being applied to the sample. 