View Featured Offers >>

SignalStar™ Secondary Antibody Kit for Use Manually in SignalStar™ Multiplex Immunohistochemistry

Product Information:

signalstar-8plex  

Storage: Store all kit components at -20°C.

Stability: All components in this kit are stable for at least 12 months when stored at the recommended temperature. Do not exceed 5 freeze/thaw cycles.

Application: The SignalStar kits are intended for fluorescent multiplex immunohistochemistry.

Slide Number: This kit contains sufficient materials for the staining of 10 slides.

SignalStar Symbols

Contents


1 Introduction: The SignalStar™ Staining Methodology


SignalStar™ multiplex Immunohistochemistry (mIHC) is a technology that employs antibodies, oligonucleotides (oligos), and fluorophores to interrogate the cellular presence, location, function, and biomarker co-expression patterns. SignalStar technology enables the detection of multiple phenotypic and functional targets while maintaining spatial context and tissue architecture. These insights are essential for understanding how cells organize and interact to influence the tissue microenvironment and drive disease progression and response to therapy.

The power of the SignalStar system lies in the design of the SignalStar antibodies. These antibodies have been rigorously validated for use in formalin-fixed, paraffin-embedded (FFPE) tissues, and subsequently conjugated to unique oligo tags using site-specific conjugation and thorough purification methodologies. Using a highly specific network of complementary oligos and fluorophores, scientists can amplify the signal for 3-8 targets, even if they are in low abundance.

SignalStar Multiplex IHC WorkFlow Diagram

Figure 1. All antibodies in your plex size of choice (3-8 maximum unique oligo-conjugated antibodies) are added in cocktail in one primary incubation step. Complementary oligos with fluorescent dyes (channels: 488, 594, 647, and 750) amplify the signal of up to 4 oligo-conjugated antibodies in the first round of imaging by building oligo-fluorophore constructs attached to the antibody. If the plex size is greater than 4, the first round of oligos and fluorophores are gently removed, and a second round of amplification is performed to visualize up to 4 additional oligo-conjugated antibodies; the complementary oligo system and the use of the fluorophore removal process enables a second round of antibodies to be amplified from the same substrate, without cross-reactivity. The 2 images are then aligned and fused computationally with either proprietary or open-source software to generate an image consisting of up to 8 targets.

2 Solutions and Reagents


2.1 Included Kit Components

Materials Included in SignalStar™ Secondary Antibody Kit
Up to 3 SignalStar oligo-conjugated secondary antibodies (see below)
Up to 3 SignalStar complementary oligos (see below)
Up to 3 Secondary blocking isotype control antibody
SignalStar™ Antibody Diluent A
SignalStar™ Antibody Diluent B
Materials Included in SignalStar™ Multiplex IHC Buffer Kit
Up to 7 SignalStar oligo-conjugated antibodies (see below)
Up to 7 SignalStar complementary oligos (see below)
SignalStar™ Antibody Diluent A
SignalStar™ Antibody Diluent B
SignalStar™ Amplification Buffer A
SignalStar™ Amplification Buffer B
SignalStar™ Amplification Oligo Set A:
 488
 594
 647
 750
SignalStar™ Amplification Oligo Set B:
 488
 594
 647
 750
SignalStar™ Ligation Buffer
T4 DNA Ligase (5 U/µL)
ATP (100 mM)
10X dsDNase Buffer
dsDNase
 

Included are SignalStar oligo-conjugated antibodies (A) and SignalStar complementary oligos (B) selected at the time of order and provided in sleeves with their respective oligo-antibody pair (C). See example below.

SignalStar Oligo Antibodies

2.2 Required Reagents Not Included

  • IMPORTANT: Primary antibody required for SignalStar secondary antibody detection
  • IMPORTANT: SignalStain® Antibody Diluent #8112, or other recommended primary antibody diluent
  • SignalStain® EDTA Unmasking Solution (10X) #14747
  • Xylene (for deparaffinization)
  • Ethanol, anhydrous denatured, histological grade (100% and 95%)
  • Decloaking Chamber (Biocare Medical, #DC2012)
  • Tris Buffered Saline with Tween 20 (TBST-10X) #9997
  • DAPI #4083
  • ProLong Gold Antifade Reagent #9071
  • Nuclease-free Water #12931
  • 10% Neutral Buffered Formalin
  • Low Retention Pipette Tips
  • Slide Processing Containers
  • Hydrophobic Barrier Pen
  • Glass Coverslips
  • Charged Slides
  • Control Tissues

3 Important Considerations Before You Begin


Warning

PLEASE READ SOLUTION PREPARATION AND PROTOCOL IN ITS ENTIRETY PRIOR TO SETTING UP YOUR EXPERIMENT.

Warning

DO NOT COMBINE COMPLEMENTARY OLIGOS OF THE SAME FLUORESCENT CHANNEL. Imaging rounds can contain only 1 complementary oligo for each fluorescent channel. DO NOT COMBINE COMPLEMENTARY OLIGOS SPECIFIC TO THE SAME FLUORESCENT CHANNEL IN THE SAME IMAGING ROUND, AS IT WILL RENDER THE ASSAY RESULTS UNINTERPRETABLE.

 
Warning

DO NOT COMBINE ANTIBODIES FROM DIFFERENT PANELS. If you are running multiple panels simultaneously, a separate antibody/complementary oligo mix must be generated for each unique panel.

Warning

SOME SIGNALSTAR KIT COMPONENTS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

Caution

Please confirm whether your SignalStar panel design requires 2 rounds of imaging. Utilize the Example SignalStar Panel Design and SignalStar Panel Design Worksheet in section 9.1 and 8.1 of this document for guidance and assistance.

Caution

Slides should be imaged within 8 hr of staining completion. Fluorescent signal may be diminished if slides are not imaged within this timeframe.

Caution

Confirm your microscope can detect the fluorophores provided in this kit. When imaging, there are 4 fluorescent channels in addition to DAPI that need to be acquired. Contact your instrument provider to confirm instrument settings.

 
Fluorophore Channel Excitation (nm) Emission (nm) Laser Line Common Filter Set
488 488 520 488 FITC
594 590 618 561/594 Texas Red
647 650 668 594/633 Cy®5
750 752 776 633 Cy®7
Note

It is recommended that a spectral library be created by imaging single stained slides for the spectra described above. This can enable better channel separation to help minimize the possibility of spectral bleed-through.

Note

Usage of a DAPI concentrate is recommended rather than a mount that contains DAPI. Bright DAPI staining facilitates better image alignment.

Note

Results are not guaranteed if there is any deviation from this protocol. The SignalStar protocol was developed and optimized with the designated antigen retrieval and staining steps.

Note

The usage of a positive control slide is recommended. A tissue upon which chromogenic staining has confirmed the presence of all targets in the multiplex panel should be included in each run.

Note

Tissue preparation and fixation methods may vary, therefore enough antibody reagent is supplied for use at a range of 1:50 to 1:200 dilution. It recommended that each antibody be used at 1:100 for initial testing to determine whether titration is required.

Note

Drain off the incubation solutions and dH2O as much as possible throughout the staining process without allowing slides to dry out. Thoroughly flick liquid off from every slide after each step before continuing to the next.

Note

The following SignalStar kit components can be thawed overnight at 4C prior to use:

  • SignalStar™ Antibody Diluent A
  • SignalStar™ Antibody Diluent B
  • SignalStar™ Amplification Buffer A
  • SignalStar™ Amplification Buffer B
  • SignalStar™ Ligation Buffer

 

4 SignalStar Imaging Round 1: Solution Preparation


Warning

EACH SOLUTION SHOULD BE CREATED FRESH AND USED PROMPTLY. Please read SignalStar Imaging Round 1: Protocol for Use in section 5 in its entirety prior to creating solutions.

Caution

SignalStar kit components should be thawed at room temperature immediately prior to use unless otherwise indicated, and then stored on ice while in use.

4.1 Primary Antibody Solution

Caution

Primary antibody diluent and primary antibodies are NOT included in this kit. Please use primary antibodies at recommended dilutions provided by the manufacturer in the recommended diluent. Confirm your final volumes match what is listed in the table below.

Caution

When purchasing your primary antibody, please make sure your primary antibody is validated for use with formalin-fixed, paraffin embedded tissues. You can see a comprehensive catalog of IHC-P-validated antibodies at www.cellsignal.com.

Warning

IF USING MORE THAN ONE PRIMARY AND SECONDARY ANTIBODY: Combine all primary antibodies from unique source species (ex. Rabbit mAb, Mouse mAb, and Rat mAb). DO NOT COMBINE PRIMARY ANTIBODIES FROM THE SAME SOURCE SPECIES.

 

Kit Component / Reagent
Volume Required
5 Slides (µL) 10 Slides (µL)
Primary Antibody Diluent (ex. SignalStain® Antibody Diluent #8112)    
Each Primary Antibody (ex. FoxP3 (D2W8ETM) Rabbit mAb #98377)    
Total Volume 750 1,500

4.2 SignalStar Secondary Antibody Solution

Warning

SIGNALSTAR ANTIBODY DILUENTS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

Caution

IF USING MORE THAN ONE PRIMARY AND SECONDARY ANTIBODY: Combine all secondary antibodies and complementary oligos used to detect unique source species (ex. Anti-rabbit, Anti-mouse, and Anti-rat)..

 

Kit Component / Reagent
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Antibody Diluent A 525 1,050
SignalStar Antibody Diluent B 225 450
Each Secondary Antibody - SignalStar Conjugate 7.5 15
Each Secondary Antibody - Complementary Oligo 7.5 15
Total Volume Up to 795 Up to 1,590

4.3 SignalStar Secondary Blocking Solution

Warning

SIGNALSTAR ANTIBODY DILUENTS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

Caution

IF USING MORE THAN ONE PRIMARY AND SECONDARY ANTIBODY: Combine all secondary blocking isotype control antibodies for included unique source species (ex. Rabbit mAb, Mouse mAb, and Rat mAb)..

 

Kit Component / Reagent
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Antibody Diluent A 525 1,050
SignalStar Antibody Diluent B 225 450
Each Secondary Blocking Isotype Control 7.5 15
Total Volume Up to 772.5 Up to 1,545

4.4 Imaging Round 1 Solution

Caution

Once prepared, the SignalStar Imaging Round 1 Solution should contain ALL antibodies (up to 7) ordered with your kit (including those for Imaging Round 1 and Imaging Round 2) and the complementary oligos for Imaging Round 1. Utilize the Example SignalStar Panel Design and SignalStar Panel Design Worksheet in section 9.1 and 8.1 of this document for guidance and assistance.

Warning

DO NOT COMBINE COMPLEMENTARY OLIGOS OF THE SAME FLUORESCENT CHANNEL. Do not combine complementary oligos from Imaging Round 1 and Imaging Round 2.

Warning

SIGNALSTAR ANTIBODY DILUENTS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Component
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Antibody Diluent A 525 1,050
SignalStar Antibody Diluent B 225 450
Each SignalStar Oligo-Conjugated Antibody 7.5 15
Each Imaging Round 1: SignalStar Complementary Oligo 7.5 15
Total Volume Up to 825 Up to 1,650

4.5 Imaging Round 1: Amplification Solution 1

Warning

SIGNALSTAR AMPLIFICATION BUFFERS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Component
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Amplification Buffer A 3,000 6,000
SignalStar Amplification Buffer B 3,000 6,000
SignalStar Amplification Oligo A-488 60 120
SignalStar Amplification Oligo A-594 60 120
SignalStar Amplification Oligo A-647 60 120
SignalStar Amplification Oligo A-750 60 120
Total Volume 6,240 12,480

4.6 Imaging Round 1: Amplification Solution 2

Warning

SIGNALSTAR AMPLIFICATION BUFFERS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Component
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Amplification Buffer A 3,000 6,000
SignalStar Amplification Buffer B 3,000 6,000
SignalStar Amplification Oligo B-488 60 120
SignalStar Amplification Oligo B-594 60 120
SignalStar Amplification Oligo B-647 60 120
SignalStar Amplification Oligo B-750 60 120
Total Volume 6,240 12,480

4.7 Imaging Round 1: Ligation Solution

Caution

Store all SignalStar kit components on ice when preparing solutions. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Components/Reagents
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Ligation Buffer 375 750
Nuclease-free Water #12931 352.5 705
T4 DNA Ligase (5 U/µL) 15 30
ATP (100 mM) 7.5 15
Total Volume 750 1,500

4.8 Additional Required Solutions Not Included

  • 1X EDTA Unmasking Solution: To prepare 250 mL of 1X EDTA Unmasking Solution, dilute 25 mL of SignalStain® EDTA Unmasking Solution (10X) #14747 with 225 mL of dH2O.
  • 10% Neutral Buffered Formalin
  • 1X Tris Buffered Saline with Tween 20 (TBST): To prepare 1 L 1X TBST, add 10 mL Tris Buffered Saline with Tween 20 (TBST-10X) #9997 to 900 mL of dH2O, mix.

5 SignalStar Imaging Round 1: Protocol for Use


Warning

EACH SOLUTION SHOULD BE CREATED FRESH AND USED PROMPTLY. Please read SignalStar Imaging Round 1: Protocol for Use in section 5 in its entirety prior to creating solutions.

5.1 Slide Baking

Note

Slide baking can be performed the day before you begin your experiment. This step allows for the paraffin wax to melt and the tissue better adhere to the slide.

 

1. Incubate slides for at least 30 min at 60°C.

5.2 Deparaffinization/Hydration

Caution

Do not let slides dry out at any point once they are deparaffinized. Use a humidified chamber for all incubation steps. Make sure each solution covers the entirety of the tissue.

 

2. Incubate sections in 3 washes of xylene for 5 min each.

 

3. Incubate sections in 2 washes of 100% ethanol for 10 min each.

 

4. Incubate sections in 2 washes of 95% ethanol for 10 min each.

 

5. Wash sections 2 times in dH2O for 5 min each.

5.3 Antigen Unmasking

Note

EDTA antigen retrieval in a pressure cooker is recommended to maximize retrieval of epitopes. This protocol describes the conditions that are recommended for the Biocare Medical Decloaking Chamber #DC2012. Device-specific settings. Manufacturer recommended operating instructions should be utilized for other pressure cookers.

 
6. 
 
Place slides in 250 mL room temperature 1X EDTA Unmasking Solution in a 24-slide holder. Add blank slides to completely fill the slide holder.
 

7. Place 500 mL dH2O into the pressure cooker.

 

8. Place the slide holder into the pressure cooker, touching the heat shield. Partially cover with a slide container lid.

Note

It may be advantageous to place a second 24-slide holder filled with 250 mL water and blank slides into the pressure cooker, and partially cover with a lid.

 

9. Seal the chamber and proceed with retrieval. Settings for the Biocare Medical Decloaking Chamber #DC2012 are 110°C for 30 min.

 

10. Carefully vent the device, then remove the lid.

 

11. Remove the slide container from the decloaking chamber and allow to cool on the bench top for 10 min.

 
12. 
 
Place the container of slides under a dH2O faucet and add water directly to the slide container until all EDTA is replaced by dH2O.

5.4 Primary and Secondary Antibody Staining

 

13. Thoroughly flick off liquid from slides and incubate in 150 µL Primary Antibody Solution according to manufacturer's protocol.

Note

Slides can be incubated overnight in the Primary Antibody Solution at 4°C in order to break up the protocol into multiple days.

 

14. Thoroughly flick off liquid from slides and immerse in 1X TBST for 30 sec.

 

15. Thoroughly flick off liquid from the slide and incubate in 150 µL Secondary Antibody Solution for 40 min at room temperature.

 

16. Thoroughly flick off liquid from slides and immerse in 1X TBST for 30 sec.

 

17. Thoroughly flick off liquid from slide and incubate in 150 µL Secondary Blocking Solution for 40 min at room temperature.

 

18. Thoroughly flick off liquid from slides and immerse in 1X TBST for 30 sec.

 

19. Incubate slides in 10% Neutral Buffered Formalin for 5 min at room temperature.

 

20. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

5.5 Imaging Round 1: Staining

 

21. Incubate slides in 150 µL of SignalStar Imaging Round 1 Solution for 40 min at room temperature.

Note

Slides can be incubated overnight at 4°C in order to break up the protocol into multiple days.

 

22. Thoroughly flick off liquid from slides and immerse in 1X TBST for 30 sec.

 

23. Incubate slides in 10% Neutral Buffered Formalin for 5 min at room temperature.

 

24. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

5.6 Imaging Round 1: Amplification

 
25. 
 
Perform 8 rounds of amplification with Amplification Solutions 1 and 2. Use the SignalStar Imaging Round 1 Checklist in section 8.2 to track your progress. Each amplification round consists of the following steps:
  1. Incubate slides in 150 µL of Amplification Solution 1 for 8 min.
  2. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.
  3. Incubate slides in 150 µL of Amplification Solution 2 for 8 min.
  4. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

5.7 Imaging Round 1: Ligation

 

26. Incubate slides in 150 µL of SignalStar Ligation Buffer for 20 min at room temperature.

 

27. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

5.8 Imaging Round 1: Image

 

28. Prepare DAPI #4083 solution according to datasheet instructions.

 

29. Immerse slides in 1X TBST for 30 sec.

 

30. Counterstain with DAPI solution.

 

31. Immerse slides in 1X TBST for 30 sec.

 

32. Mount slides with ProLong Gold Antifade Reagent #9071.

 

33. Image slides as soon as possible. Signal should remain robust for up to 8 hr.

Note

Once imaging is complete, fluorescent signal can be removed from slides so that another round of imaging can be performed. Perform Imaging Round 2 as soon as possible following the first round of imaging.

6 SignalStar Imaging Round 2: Solution Preparation


Warning

EACH SOLUTION SHOULD BE CREATED FRESH AND USED PROMPTLY. Please read SignalStar Imaging Round 2: Protocol for Use in section 7 in its entirety prior to creating solutions.

Caution

SIGNALSTAR IMAGING ROUND 2 IS ONLY NEEDED FOR ANTIBODY PANELS THAT REQUIRE TWO IMAGING ROUNDS. Utilize the Example SignalStar Panel Design and SignalStar Panel Design Worksheet in section 9.1 and 8.1 of this document for guidance and assistance.

Caution

SignalStar kit components should be thawed at room temperature immediately prior to use unless otherwise indicated, and then stored on ice while in use.

6.1 Fluorescence Removal Solution

Caution

Store all SignalStar kit components on ice when preparing solutions. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Components/Reagents
Volume Required
5 Slides (µL) 10 Slides (µL)
10X dsDNase Buffer 75 150
dsDNase 7.5 15
Nuclease-free Water #12931 667.5 1,335
Total Volume 750 1,500

6.2 Imaging Round 2 Solution

Caution

Once prepared, the SignalStar Imaging Round 2 Solution should contain only the complementary oligos for Imaging Round 2. Utilize the Example SignalStar Panel Design and SignalStar Panel Design Worksheet in section 9.1 and 8.1 of this document for guidance and assistance.

Warning

NO SIGNALSTAR CONJUGATES ARE ADDED TO THE IMAGING ROUND 2 SOLUTION.

Warning

DO NOT COMBINE COMPLEMENTARY OLIGOS OF THE SAME FLUORESCENT CHANNEL. Do not combine complementary oligos from Imaging Round 1 and Imaging Round 2.

Warning

SIGNALSTAR ANTIBODY DILUENTS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Components
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Antibody Diluent A 525 1,050
SignalStar Antibody Diluent B 225 450
Each Imaging Round 2: SignalStar Complementary Oligo 7.5 15
Total Volume Up to 780 Up to 1,560

6.3 Imaging Round 2: Amplification Solution 1

Warning

SIGNALSTAR ANTIBODY DILUENTS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Components
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Amplification Buffer A 3,000 6,000
SignalStar Amplification Buffer B 3,000 6,000
SignalStar Amplification Oligo A-488 60 120
SignalStar Amplification Oligo A-594 60 120
SignalStar Amplification Oligo A-647 60 120
SignalStar Amplification Oligo A-750 60 120
Total Volume 6,240 12,480

6.4 Imaging Round 2: Amplification Solution 2

Warning

SIGNALSTAR ANTIBODY DILUENTS ARE VISCOUS. FLUORESCENT SIGNAL MAY BE VARIABLE OR DIMINISHED IF SOLUTIONS ARE NOT ACCURATELY MEASURED OR SUFFICIENTLY MIXED. Combine SignalStar kit components in 15 mL conical tubes using low retention pipette tips. Pipette slowly to ensure accuracy. Rotate end-over-end for 20 min at room temperature. Store all SignalStar kit components on ice when not in use. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Components
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Amplification Buffer A 3,000 6,000
SignalStar Amplification Buffer B 3,000 6,000
SignalStar Amplification Oligo B-488 60 120
SignalStar Amplification Oligo B-594 60 120
SignalStar Amplification Oligo B-647 60 120
SignalStar Amplification Oligo B-750 60 120
Total Volume 6,240 12,480

6.5 Imaging Round 2: SignalStar Ligation Solution

Caution

Store all SignalStar kit components on ice when preparing solutions. Once combined, SignalStar solutions should be kept at room temperature and used promptly.

 

Kit Components/Reagents
Volume Required
5 Slides (µL) 10 Slides (µL)
SignalStar Ligation Buffer 375 750
Nuclease-free Water #12931 352.5 705
T4 DNA Ligase (5 U/µL) 15 30
ATP (100 mM) 7.5 15
Total Volume 750 1,500

7 SignalStar Imaging Round 2: Protocol for Use


Warning

EACH SOLUTION SHOULD BE CREATED FRESH AND USED PROMPTLY. Please read SignalStar Imaging Round 2: Protocol for Use in section 7 in its entirety prior to creating solutions.

Caution

SIGNALSTAR IMAGING ROUND 2 IS ONLY NEEDED FOR ANTIBODY PANELS THAT REQUIRE TWO IMAGING ROUNDS. Utilize the Example SignalStar Panel Design and SignalStar Panel Design Worksheet in section 9.1 and 8.1 of this document for guidance and assistance.

7.1 Removal of Fluorescent Signal

 

1. After image acquisition for Imaging Round 1, soak slides in dH2O for ≥30 min to gently remove coverslips without damaging tissue.

 

2. Incubate slides in 150 µL of Fluorescence Removal Solution for 2 hr at 37°C.

 

3. Immerse slides in dH2O for 30 sec.

 

4. Optional:

  1. Mount slides in ProLong Gold Antifade Reagent #9071 with DAPI #4083 and image slides to confirm that no fluorescent signal remains.
  2. Soak slides in dH2O for ≥ 30 min to gently remove coverslips without damaging tissue.
  3. Immerse slides in dH2O for 30 sec.

7.2 Imaging Round 2: Staining

 

5. Incubate slides in 150 µL of SignalStar Imaging Round 2 Solution for 40 min at room temperature.

 

6. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

7.3 Imaging Round 2: Amplification

 
7. 
 
Perform 8 rounds of amplification with Amplification Solutions 1 and 2. Use the SignalStar Imaging Round 2 Checklist in section 8.3 to track your progress. Each amplification round consists of the following steps:
  1. Incubate slides in 150 µL of Amplification Solution 1 for 8 min.
  2. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.
  3. Incubate slides in 150 µL of Amplification Solution 2 for 8 min.
  4. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

7.4 Imaging Round 2: Ligation

 

8. Incubate slides in 150 µL of SignalStar Ligation Buffer for 20 min at room temperature.

 

9. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

7.5 Imaging Round 2: Image

 

10. Prepare DAPI #4083 solution according to datasheet instructions.

 

11. Immerse slides in 1X TBST for 30 sec.

 

12. Counterstain with DAPI solution.

 

13. Immerse slides in 1X TBST for 30 sec.

 

14. Thoroughly flick off liquid from slides and immerse in dH2O for 30 sec.

 

15. Mount slides with ProLong Gold Antifade Reagent #9071.

 

16. Image slides as soon as possible. Signal should remain robust for up to 8 hr.

8 User Worksheets


8.1 SignalStar Panel Design Worksheet

Oligo-Conjugated Antibody Complementary Oligo Product Pair # Imaging Round 488 594 647 750
               
               
               
               
               
               
               
               

(See Appendix 9.1 for an example panel design.)

8.2 SignalStar Imaging Round 1 Checklist

Amplification Round Step # Step
Amplification Round 1 1 Incubate in Amplification Solution 1 for 8 min.
2 Immerse slides in dH2O for 30 sec.
3 Incubate in Amplification Solution 2 for 8 min.
4 Immerse slides in dH2O for 30 sec.
Amplification Round 2 5 Incubate in Amplification Solution 1 for 8 min.
6 Immerse slides in dH2O for 30 sec.
7 Incubate in Amplification Solution 2 for 8 min.
8 Immerse slides in dH2O for 30 sec.
Amplification Round 3 9 Incubate in Amplification Solution 1 for 8 min.
10 Immerse slides in dH2O for 30 sec.
11 Incubate in Amplification Solution 2 for 8 min.
12 Immerse slides in dH2O for 30 sec.
Amplification Round 4 13 Incubate in Amplification Solution 1 for 8 min.
14 Immerse slides in dH2O for 30 sec.
15 Incubate in Amplification Solution 2 for 8 min.
16 Immerse slides in dH2O for 30 sec.
Amplification Round 5 17 Incubate in Amplification Solution 1 for 8 min.
18 Immerse slides in dH2O for 30 sec.
19 Incubate in Amplification Solution 2 for 8 min.
20 Immerse slides in dH2O for 30 sec.
Amplification Round 6 21 Incubate in Amplification Solution 1 for 8 min.
22 Immerse slides in dH2O for 30 sec.
23 Incubate in Amplification Solution 2 for 8 min.
24 Immerse slides in dH2O for 30 sec.
Amplification Round 7 25 Incubate in Amplification Solution 1 for 8 min.
26 Immerse slides in dH2O for 30 sec.
27 Incubate in Amplification Solution 2 for 8 min.
28 Immerse slides in dH2O for 30 sec.
Amplification Round 8 29 Incubate in Amplification Solution 1 for 8 min.
30 Immerse slides in dH2O for 30 sec.
31 Incubate in Amplification Solution 2 for 8 min.
32 Immerse slides in dH2O for 30 sec.

8.3 SignalStar Imaging Round 2 Checklist

Amplification Round Step # Step
Amplification Round 1 1 Incubate in Amplification Solution 1 for 8 min.
2 Immerse slides in dH2O for 30 sec.
3 Incubate in Amplification Solution 2 for 8 min.
4 Immerse slides in dH2O for 30 sec.
Amplification Round 2 5 Incubate in Amplification Solution 1 for 8 min.
6 Immerse slides in dH2O for 30 sec.
7 Incubate in Amplification Solution 2 for 8 min.
8 Immerse slides in dH2O for 30 sec.
Amplification Round 3 9 Incubate in Amplification Solution 1 for 8 min.
10 Immerse slides in dH2O for 30 sec.
11 Incubate in Amplification Solution 2 for 8 min.
12 Immerse slides in dH2O for 30 sec.
Amplification Round 4 13 Incubate in Amplification Solution 1 for 8 min.
14 Immerse slides in dH2O for 30 sec.
15 Incubate in Amplification Solution 2 for 8 min.
16 Immerse slides in dH2O for 30 sec.
Amplification Round 5 17 Incubate in Amplification Solution 1 for 8 min.
18 Immerse slides in dH2O for 30 sec.
19 Incubate in Amplification Solution 2 for 8 min.
20 Immerse slides in dH2O for 30 sec.
Amplification Round 6 21 Incubate in Amplification Solution 1 for 8 min.
22 Immerse slides in dH2O for 30 sec.
23 Incubate in Amplification Solution 2 for 8 min.
24 Immerse slides in dH2O for 30 sec.
Amplification Round 7 25 Incubate in Amplification Solution 1 for 8 min.
26 Immerse slides in dH2O for 30 sec.
27 Incubate in Amplification Solution 2 for 8 min.
28 Immerse slides in dH2O for 30 sec.
Amplification Round 8 29 Incubate in Amplification Solution 1 for 8 min.
30 Immerse slides in dH2O for 30 sec.
31 Incubate in Amplification Solution 2 for 8 min.
32 Immerse slides in dH2O for 30 sec.

9 Appendix


9.1 Example SignalStar Panel Design

Oligo-Conjugated Antibody Complementary Oligo Product Pair # Imaging Round 488 594 647 750
Goat Anti-Rabbit IgG (H&L) Antibody Complementary Oligo
(CO-0055-488)
30599 1      
PD-L1 (E1L3N®) XP® Rabbit mAb
(SignalStar™ Conjugate 0005)
Complementary Oligo
(CO-0005-594)
28249 1      
TIM-3 (D5D5R™) XP® Rabbit mAb
(SignalStar™ Conjugate 0010)
Complementary Oligo
(CO-0010-647)
15231 1      
Ki-67 (8D5) Mouse mAb
(SignalStar™ Conjugate 0014)
Complementary Oligo
(CO-0014-750)
56398 1      
CD8ɑ (D8A8Y) Rabbit mAb
(SignalStar™ Conjugate 0004)
Complementary Oligo
(CO-0004-488)
45747 2      
CD68 (D4B9C) XP® Rabbit mAb
(SignalStar™ Conjugate 0007)
Complementary Oligo
(CO-0007-594)
77318 2      
CD20 (E7B7T) XP® Rabbit mAb
(SignalStar™ Conjugate 0011)
Complementary Oligo
(CO-0011-647)
36775 2      
Pan-Keratin (C11) Mouse mAb
(SignalStar™ Conjugate 0003)
Complementary Oligo
(CO-0003-750)
97227 2      

9.2 Troubleshooting and Frequently Asked Questions

How are the SignalStar Multiplex IHC Kits & Reagents validated?

CST thoroughly validates each antibody available in the SignalStar Multiplex IHC Panel Builder menu. Various combinations of antibodies are tested through titration and fluorophore pairing, and in both rounds of imaging. Testing is performed on a variety of tumors and tissue types. We also rigorously test the parent antibodies used in the traditional chromogenic assay, as they serve as the foundation of this fluorescent assay. We are not able to test all multiplex configurations or tissues. Please contact customer support if you have any questions.

Do I need to optimize the SignalStar Multiplex IHC Kits & Reagents for the type of tissue I’m using?

The SignalStar Multiplex IHC Kits & Reagents have been optimized with respect to fluorophore pairing and order of antibodies. As tissues vary in quality and expression level of targets, increasing the concentration of antibodies in your panel by 2-fold or decreasing by 0.5 fold can help achieve optimal signal in your experiments. Small titrations such as this may be necessary in order to achieve lower levels of background and higher S/N levels.

What are the imaging instrument specifications required to image SignalStar Multiplex IHC?

There are four fluorescent channels in addition to DAPI that need to be acquired. Contact your instrument provider to confirm instrument specifications. Ensure the Texas Red filter set is used to visualize the 594 channel. The TRITC filter is not ideal and may result in diminished signal output.

Fluorophore Channel Excitation (nm) Emission (nm) Laser Line Common Filter Set
488 488 520 488 FITC
594 590 618 561/594 Texas Red
647 650 668 594/633 Cy®5
750 752 776 633 Cy®7

SignalStar mIHC is a signal amplification-based technology, therefore for initial imaging, we recommend auto-exposing control tissues to identify optimal exposure times for each channel.

For some instruments, spectral bleed-through can occur between the 594 and 647 channels. Decreasing the concentration of antibodies placed in the 647 channel may help with bleed-through. Spectral bleed-through can also be considered during panel design so that a strong phenotypic marker is spectrally separated from weaker expressing targets.

Does this assay work on frozen tissue?

SignalStar Multiplex IHC Kits & Reagents haven’t yet been validated for use in frozen tissues.

Do you have mouse-reactive antibodies available?

Yes, please select "Mouse" in the first step of the SignalStar Multiplex IHC Panel Builder to see our mouse-reactive menu.

I don't see my target of interest in your menu of available antibodies. Can I still use it in my panel in some way?

SignalStar Multiplex IHC Kits & Reagents haven't yet been validated for use with antibodies outside of our menu. We're in the process of developing custom solutions for using your own antibodies in the SignalStar Multiplex IHC assay.

SignalStar mIHC is a non-destructive technology. Therefore, in order to use your antibodies of interest, you may perform direct immunofluorescence on the same tissue after the SignalStar assay. The SignalStar™ Fluorescence Removal Kit #32722 enables you to remove the fluorescent oligos after SignalStar mIHC in order to stain and visualize direct immunofluorescence.

Can I combine antibodies used in this assay with direct conjugates?

SignalStar Multiplex IHC Kits & Reagents have been validated for use in combination with direct conjugates. The SignalStar™ Fluorescence Removal Kit #32722 enables you to remove the fluorescent oligos after SignalStar mIHC in order to stain and visualize direct immunofluorescence directly after SignalStar mIHC. We have found that many direct conjugates against strong cell surface markers work well when used in this manner. Please see our recent poster presented at AACR this past spring for more information. In addition, please see our line of directly conjugated antibodies for use in direct immunofluorescence imaging.

When comparing my SignalStar staining to the chromogenic staining on serial sections, I see more positive cells. How do I know if this excess staining is correct?

During the course of optimization, we've found that fluorescent staining may show higher %-positivity than chromogenic staining. To ensure any excess staining is specific, confirm that the correct subcellular localization and co-localization with other stains are demonstrated. For example, if all CD8+ cells are CD3+, any excess CD8+ staining compared to the chromogenic is most likely correct.

How long after the completion of staining can I wait to image my slides?

For Imaging Round 1, the staining should show robust signal when imaged up to 8 hr post completion of staining. For Imaging Round 2, imaging should be performed as close to the completion of staining as possible, but should remain robust for up to 8 hr.

Do I need to optimize the SignalStar Multiplex IHC Kits & Reagents for the type of tissue I'm using?

The SignalStar Multiplex IHC Kits & Reagents have been optimized with respect to fluorophore pairing and order of antibodies. As tissues vary in quality and expression level of targets, increasing the concentration of antibodies in your panel by 2-fold or decreasing by 0.5 fold can help achieve optimal signal in your experiments.

What is an appropriate positive control to include in this assay? Are multiple controls necessary?

Any tissue shown to be positive for each target via chromogenic IHC can serve as a positive control tissue. Each target will therefore require a positive control, which may sometimes necessitate multiple controls. For optimal comparison, the sections should be as close to serial as possible.

Can I perform alternative antigen retrieval methods than those provided in the protocol?

For optimal results, we do not recommend deviating from the SignalStar protocols. However, if you do not have access to the antigen retrieval method detailed in the protocol, you may try using a microwave, which has shown to result in reduced fluorescent signal. We do not recommend using a microwave when detecting low abundance targets, and we cannot guarantee your result.

To perform antigen retrieve using a microwave:

Prepare 1X EDTA Unmasking Solution: To prepare 250 mL of 1X EDTA unmasking solution, dilute 25 ml of SignalStain® EDTA Unmasking Solution (10X) (#14747) with 225 mL of dH2O. Heat slides in a microwave submersed in 1X EDTA unmasking solution until boiling is initiated (~2.5 minutes at power level 10). Follow with 15 min at a sub-boiling temperature (95°-98°C). In most common microwaves, this equates to 8 minutes at power level 3, then 7 minutes power level 2. Then remove from the microwave and place the container of slides under a dH2O faucet and add water directly to the slide container until all EDTA is replaced by dH2O. No cooling is necessary.

I am looking to run this assay on adipose tissue / brain / normal kidney which has a high level of autofluorescence. Do you have any example data that you can provide me to demonstrate that this assay will work in this tissue type?

While we utilize a wide variety of tumor and tissue types during the course of our optimization process, we cannot account for all tissues and expression levels. This assay should work in FFPE tissue 4-5 uM in thickness, assuming that our recommended protocol is followed. Furthermore, due to the high level of amplification that this assay provides, even very high autofluorescence levels may be overcome with the resulting strong, specific signal.

Can I use an alternative mounting reagent instead of ProLong Gold Antifade Reagent #9071?

Prolong Antifade mounting media are optimal for this protocol. Internal testing has demonstrated that SlowFade Antifade Reagents can negatively impact signal intensity.

Cell Signaling Technology, XP, and SignalStar are trademarks of Cell Signaling Technology, Inc.
E1L3N is a registered trademark of Cell Signaling Technology, Inc. Cy and CyDye are registered trademarks of GE Healthcare. U.S. Patent No. 10,781,477, foreign equivalents, and child patents deriving therefrom. All other trademarks are the property of their respective owners. Visit our Trademark Information page.
© 2023 Cell Signaling Technology, Inc. All Rights Reserved.